목표농도주입법에서 프로포폴 농도를 잘못 설정 후 교정 방법

DC Field Value Language
dc.contributor.advisor채윤정-
dc.contributor.authorLee, Won Il-
dc.date.accessioned2019-10-21T07:24:20Z-
dc.date.available2019-10-21T07:24:20Z-
dc.date.issued2015-02-
dc.identifier.other19017-
dc.identifier.urihttps://dspace.ajou.ac.kr/handle/2018.oak/18602-
dc.description학위논문(석사)--아주대학교 일반대학원 :의학과,2015. 2-
dc.description.abstractBackground: We investigated the correction methods following wrong-settings of emulsion concentrations of propofol as a countermeasure against erroneous target-controlled infusions (TCI). Methods: TCIs were started with targeting 4.0 μg/ml of effect-site concentration (Ceff) of propofol, and the emulsion concentrations were selected for 2.0% instead of 1.0% (FALSE1- 2, n = 24), or 1.0% instead of 2.0% (FALSE2-1, n = 24). These wrong TCIs were corrected at 3 min after infusion start. During FALSE1-2, the deficit was filled up while injecting after equilibrium (n = 12), or while overriding (n = 12). During FALSE2-1, the overdose was evacuated while targeting Ceff (n = 12) or targeting plasma concentration (Cp) (n = 12). The gravimetrical measurements of TCI reproduced the Cp and Ceff using simulations. The reproduced Ceff at 3 min (Ceff-3min) and the time to be normalized within ± 5% of target Ceff (T±5%), were compared between the correction methods. Results: During the wrong TCI, Ceff-3min was 1.98 ± 0.01 μg/ml in FALSE1-2, and 7.99 ± 0.05 μg/ml in FALSE2-1. In FALSE1-2, T±5% was significantly shorter when corrected while overriding (3.9 ± 0.25 min), than corrected after equilibrium (6.9 ± 0.05 min) (P < 0.001). In FALSE2-1, T±5% was significantly shorter during targeting Cp (3.6 ± 0.04 min) than targeting Ceff (6.7 ± 0.15 min) (P < 0.001). Conclusions: The correction methods, based on the pharmacokinetic and pharmacodynamic characteristics, could effectively and rapidly normalize the wrong TCI following erroneously selections of the emulsion concentration of propofol.-
dc.description.tableofcontentsⅠ.INTRODUCTION 1 Ⅱ.MATERIALS AND METHODS 2 Ⅲ.RESULTS 7 Ⅳ.DISCUSSION 11 Ⅴ.CONCLUSION 15-
dc.language.isoeng-
dc.publisherThe Graduate School, Ajou University-
dc.rights아주대학교 논문은 저작권에 의해 보호받습니다.-
dc.title목표농도주입법에서 프로포폴 농도를 잘못 설정 후 교정 방법-
dc.title.alternativeWon Il Lee-
dc.typeThesis-
dc.contributor.affiliation아주대학교 일반대학원-
dc.contributor.alternativeNameWon Il Lee-
dc.contributor.department일반대학원 의학과-
dc.date.awarded2015. 2-
dc.description.degreeMaster-
dc.identifier.localId695533-
dc.identifier.urlhttp://dcoll.ajou.ac.kr:9080/dcollection/jsp/common/DcLoOrgPer.jsp?sItemId=000000019017-
dc.subject.keyword약물 전달 시스템-
dc.subject.keyword주입 펌프-
dc.subject.keyword정맥 주입-
dc.subject.keywordpropofol-
dc.subject.keyworddrug delivery systems-
dc.subject.keywordinfusion pumps-
dc.subject.keywordintravenous infusion-
Appears in Collections:
Graduate School of Ajou University > Department of Medicine > 3. Theses(Master)